Arcomed Volumed µVP7000 Infusion Pump 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Admedus Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00314-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2018-05-02
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The manufacturer, arcomed ag, has been informed that in some rare cases the volumed µvp7000 did not perform within the specified volumetric accuracy with pvc lines due to low door pressure. a deviation of the infused volume can have a negative impact on the patient. the pumps do perform an initial stop flow test that also tests the door pressure, however in rare occasions pumps would still not perform within the specified tolerances after passing the stop flow test. admedus australia, following consultation with the manufacturer, have developed a 'strip test" to be included in the standard preventative maintenance testing procedures, which are used to check that the machines are operating in accordance with specifications. the strip test will check devices for mechanical tolerances of the door which will increase the safety tolerances of the device.
  • 조치
    Admedus is advising customers to add the replacement page provided with the Customer Letter to the user manual of the device. Users should continue to monitor operation of the pumps and arrange for the pump to be checked by a trained service engineer if they notice any volumetric inaccuracies (note: small deviations within the specifications (5%) may occur). Service Technicians are being provided a replacement page for the service manual instructing how to perform the 'Strip Test'.

Device

  • 모델명 / 제조번호(시리얼번호)
    Arcomed Volumed µVP7000 Infusion PumpARTG Number: 279584(Admedus Australia - Infusion pump, general-purpose)