ARCTIC SUN 5000 Temperature Management System (220-240 V Australia) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Bard Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00534-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-07-19
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    An issue regarding the use of the arctic sun 5000 temperature management system, in excess of 2,000 system hours, without conducting preventative maintenance has been identified. this issue may result in failure of certain components, in particular, the mixing pump of the system. the mixing pump runs continuously during operation and its failure may cause the system to cease functioning as intended. therefore, resulting in no availability of cooling or rewarming therapy for the patient. to date, bard australia has not received adverse reports of injuries or diseases related to this issue.
  • 조치
    Bard Australia is updating the Operators Manual and the Service Manual for the ARCTIC SUN 5000 Temperature Management System. Customers are advised to examine their inventory for affected devices to determine their total system hours of use by following the instructions detailed in the recall letter. Customers with devices approaching or exceeding 2,000 system hours are requested to contact their local Bard representative for further information. Customers with devices that have less than 2,000 hours are reminded to address potential issues before their occurrence in order to maintain device availability. In the event of a device failure, customers are advised to utilise alternative conventional methods for patient temperature control.

Device

  • 모델명 / 제조번호(시리얼번호)
    ARCTIC SUN 5000 Temperature Management System (220-240 V Australia)Product Code: 5000-01-03ARTG Number: 150422
  • Manufacturer

Manufacturer