Arrow International 4Fg Seldinger Conversion Set with Long Sheath(Peripherally inserted central venous catheter) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Teleflex Medical Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00729-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-08-12
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The reason for this action is due to the possibility that the catheter peel-away component hub tabs may prematurely detach when the practitioner begins to peel apart the sheath body from the catheter. the component tabs extend past the opening to the sheath, so even if both tabs break, there will be some length of the sheath that would protrude from the skin. therefore, embolus of the sheath should not occur. the defect is easy to recognise and the clinician would have the choice of either continuing with the split sheath or using a new one. although the issue does not pose a significant risk to the patient, it may cause a delay of treatment when placing the catheter.
  • 조치
    Customers are advised not to use or distribute any affected stock and immediately quarantine them. Teleflex Medical will arrange collection of any affected stock from affected facilities. This action has been closed-out on 18/08/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Arrow International 4Fg Seldinger Conversion Set with Long Sheath(Peripherally inserted central venous catheter)Catalogue number: PI-01451-LSBatch Numbers: 23F14J0437, 23F14K0406, 23F14B0935, 23F14H1535, 23F15A0278, 23F15D0907ARTG Number: 148783