ARROW PICC Kits with Peel-Away Sheath Component 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Teleflex Medical Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00408-1
  • 사례 위험등급
    Class III
  • 사례 시작날짜
    2017-04-05
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Arrow international has received complaints related to peel away sheath flare for the arrow picc kits. the instructions for use included with the kits do not inform the user that a skin nick or alternate enlargement method may be required for difficult insertions. the use of a skin nick reduces the possibility of sheath flare and subsequent complications. there is a remote possibility the device may cause a delay in treatment or minor damage to the vessel wall or subcutaneous tissue.
  • 조치
    Teleflex is advising users that they may be required to use an alternate enlargement method or perform a skin nick for difficult insertions. Customers are also requested to include the supplementary card (Product Notice) with their affected stock.

Device

  • 모델명 / 제조번호(시리얼번호)
    ARROW PICC Kits with Peel-Away Sheath ComponentCatalogue Numbers: PI-01351-LS, PI-01351-SS, PI-01451-LS, PI-01451-SS, PI-01552-LS, PI-01552-SS, PR-05041, PR-05041-T, PR-05042, PR-05052, PR-05052-MWMultiple Lot NumbersARTG Numbers: 148783 and 148784
  • Manufacturer

Manufacturer