Arsenal Set-Screw, a sub-component of the Arsenal Spinal Fixation System 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Scientx Australia 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01639-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-12-22
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    For certain lots of set screw, the manufacturer has become aware of a trend specific to set screw postoperative disengagement from the screw body (set screw back-out). 16 complaints were received from october 2015 through august 2016. in the short term, there is a risk that set screw post-operative disengagement from the screw body at the end of a construct may lead to hardware failure and/or pedicle screw pull-out on adjacent levels, and may also lead to revision surgery. the long term risk is that set screw post-operative disengagement from the screw body may lead to loss of correction, as well as the non-union/non-fusion of bone (pseudoarthrosis). there is also an unrelated issue for other lots of set screws where a small proportion of lots were manufactured in an oversized condition. for this issue, testing conducted by alphatec indicates that if the set screw is successfully inserted and locked to the required tightening torque, the construct will perform appropriately.
  • 조치
    For the screw back-out issue, implanting surgeons are being informed to consider more frequent follow-up visits during the initial postoperative phase(s) after surgery. For the oversized condition issue, surgeons are advised that tightening the set screws to the prescribed tightening torque (80in-lbs) should fully mitigate the possibility of an adverse screw body splay event.

Device

  • 모델명 / 제조번호(시리얼번호)
    Arsenal Set-Screw, a sub-component of the Arsenal Spinal Fixation SystemPart Number: 47127Lot Numbers affectedFor screw back-out issue:700942For oversized condition: 7689802, 7689803, 7766301ARTG Number: 178654
  • Manufacturer

Manufacturer

  • Source
    DHTGA