Artis zee and Artis zeego systems (software versions VC14 and VC21) with the A100 generator and specific manufacturing lots of 3 focal spot x-ray tube (fluoroscopic angiographic x-ray system) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-01035-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-09-25
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens has identified a potential problem with the artis zee and artis zeego systems with the a100 generator and specific lots of a 3 focal spot x-ray tube. the “small focus” of a 3 focal spot x-ray tube may sporadically fail if the case grounding detaches from the focus head and overheating occurs. further radiation release won’t be possible. an ongoing procedure can be continued by using one of the two remaining focal spots ("large focus” or "micro focus") instead. to achieve that, the operator will need to switch over to the other focus, thus causing a short interruption of the ongoing procedure.
  • 조치
    Siemens is implementing a software update that will protect the affected component from any damage in case the grounding detaches from the focus head. In case of such a potential failure, the system will automatically switch to the "Large Focus”. This action has been closed-out on 07/06/2016.