Artis zee Floorstand, Artis zee Biplane Floorstand and Artis zee Multi Purpose systems with SW version VC21B(fluoroscopic angiographic x-ray system) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00483-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-06-05
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    A possible position sensor fault in the swivel base axis not being detected by the system software may cause the c-arm system to move faster than normal when movement is enabled by the operator. the system will not move of its own accord, movement must still be initiated by the operator, however the system can unexpectedly exceed its usual speed, potentially resulting in a collision.
  • 조치
    This issue will be remedied with software update (Patch 12). Following the installation of this update, the possible speed of the Artis system will be limited to the permissible level. This action has been closed-out on 19/08/2016.