Artiste, Oncor and Primus Digital Linear Accelerators running Control Console software version 13.0.302 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00284-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-02-28
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens has identified a risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table using the simtec auto field sequence option. the following issues will be addressed in the software update solution "auto field sequence (afs) motion protection”:- risk of collision due to automatic gantry movements in case of lateral table shifts and- unwanted automatic gantry movements in case of table configuration set to “manually” and the movement order to be “table first”.
  • 조치
    Siemens is advising affected users to be aware of the updated software solution “Auto Field Sequence (AFS) Motion Protection” which will be implemented at the Control Console of their Digital Linear Accelerator. This software will not execute a movement if a potential collision is anticipated and that users may, in such cases, perform the planned movement manually. Users are further advised that despite these safety measures, there might be scenarios which are not covered by the software. Therefore, during an auto-sequence delivery, users must be aware, at all times, of any movement of the gantry, the treatment table and the applied table offsets or overrides according to the treatment plan.

Device

  • 모델명 / 제조번호(시리얼번호)
    Artiste, Oncor and Primus Digital Linear Accelerators running Control Console software version 13.0.302ARTG Number: 165502