ARTISTE, ONCOR and PRIMUS Digital Linear Accelerators Serial numbers equal to or higher than 3094 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00168-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-02-03
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens have identified that incorrect values for the rotational collimator position may be displayed due to mechanical problems of the sensor slide. this can happen if the slide, on which the sensors used to detect the rotational collimator position are mounted, is not in its fully latched position. if the user is not aware of the incorrect values the patient can be mistreated by dose at the wrong location.Siemens were also made aware of a defective weld seam at overhead suspensions. due to such a defect the weld seam might break and the overhead suspension might fall down. this can lead to a severe injury of the patient or any other person.
  • 조치
    Siemens is undertaking hardware updates regarding the collimator position detection and improper weld seam at overhead suspensions. To keep the sensor slide in its latched position, a locking plate will be mounted to the sensor slide. This will significantly reduce the risk that the sensor slide might be mechanically misaligned and cause incorrect values being displayed for the rotational collimator position. As reinforcement, a bracket will be added to the overhead suspension to ensure the stability of the connection. This will prevent the overhead suspension from falling down.