Asserachrom Von Willebrand Factor (VWF):Collagen (CB). An in vitro diagnostic medical device (IVD). 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Diagnostica Stago Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00588-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-05-10
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Following a customer complaint, stago has investigated and confirmed that asserachrom vwf:cb lot 114710(re) gives higher results compared to the other asserachrom vwf:cb tested lots for samples with a vwf:cb level above 30%. this issue is more pronounced for samples with a vwf:cb level above 50%.Based on internal testing by stago, normal samples tested are overestimated by 26% on average (relative deviation). for example: a normal plasma at 134% could be reported at 167%.The pathology samples tested and in particular the type 1 plasmas are overestimated by 38% on average (relative deviation). for example: a type 1 plasma at 56% could be reported at 75%.The quality control (qc) included in the kit does not always detect the issue.
  • 조치
    Diagnostica Stago is advising users to stop using and destroy the affected lot according to current local regulations. Replacement kits will be provided.

Device

  • 모델명 / 제조번호(시리얼번호)
    Asserachrom Von Willebrand Factor (VWF):Collagen (CB). An in vitro diagnostic medical device (IVD). Reference Number: 00239Lot Number: 114710(RE)ARTG Number: 185170