Astral 100 and Astral150 Ventilators - Degraded Battery Pack 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Resmed Limited 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01074-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2016-08-16
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Resmed has received a small number of reports regarding the performance of the internal battery in the astral ventilator. in these reports, an internal electrical issue has led to ventilation ceasing without either the low battery alarm or the critically low battery alarm being activated. in all of these reported cases the “total power failure” alarm activated as intended.Under these circumstances, ventilator dependent patients may be at risk if no urgent action is taken to restore power or switch to back-up ventilation. the risk of this occurring is increased if the internal battery has been charged a high number of times. this fault has also been reported with the astral external battery however, there have been no reports of cessation of ventilation when the astral external battery is used.It remains safe to continue using astral ventilators, provided patients/carers follow the precautions provided. there have been no reports of adverse health effects as a result of this issue.
  • 조치
    ResMed is providing users with work around instructions to follow as an interim measure. Patients and carers are advised to connect the ventilator to AC power whenever possible, and to use an external power supply where necessary. It is recommended that ventilator dependant patients have alternative ventilation equipment available at all times. All ventilators with an affected battery pack should have their internal batteries replaced by no later than 1 December 2016. For further information, please see https://www.tga.gov.au/alert/astral-100-and-astral-150-ventilators-0