Astral Ventilators - ResMed Astral 100 & Astral 150 with SR1.1 software (SX544-0301) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Resmed Limited 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00050-2
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-01-18
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Astral devices may experience an issue when powered by an external battery (astral external battery or rps ii). the issue can prevent the user from powering on the device after the auto power off feature powers off the device when there has been no user interaction with the device for 15 minutes and the ventilator is in stand-by.The external battery start up issue only affects astral 100 and astral 150 devices with sr1.1 software (sx544-0301).
  • 조치
    ResMed is advising users to contact ResMed technical support to arrange for the installation of a software upgrade which will address the reported issue. Please note that this software upgrade (SX544-0401) addresses the circuit disconnection alarm issue reported previously (RC-2015-RN-00391-1) as well as the external battery power issue reported in this current recall notification. This action has been closed-out on 05/09/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Astral Ventilators - ResMed Astral 100 & Astral 150 with SR1.1 software (SX544-0301)Product Codes 27081 and 27083ARTG Number: 219429
  • Manufacturer

Manufacturer