Attachment for Acetabular and Medullary Reaming (used with the Compact Air Drive II, an air-driven power tool for use in traumatology, endoprosthetics and spinal column surgery) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Johnson & Johnson Medical Pty Ltd t/a DePuy Synthes 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-01094-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2015-11-10
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The geometrical shape of a subcomponent of this attachment was inadvertently changed. this change allows potential unintended reverse (counter-clockwise) motion if the subcomponent is aligned in a specific orientation when the attachment is connected to the handpiece and both triggers are pressed.This attachment is intended to be used without reverse motion. unintentional activation in reverse could potentially damage flexible drilling or reaming shafts and could potentially cause serious injury to the patient.
  • 조치
    Customer to return affected product. Sponsor to replace affected product. This action has been closed-out on 24/08/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Attachment for Acetabular and Medullary Reaming (used with the Compact Air Drive II, an air-driven power tool for use in traumatology, endoprosthetics and spinal column surgery)Product Number 511.785Serial Numbers: 16384,16385,16386,16387,16389,16390,16391,16392,16393, 16394,16395,16396,16397,16403,16405,1641ARTG Number: 157072