AU480 Clinical Chemistry Analyser - software version 1.81(Systems configured with Closed Channels ONLY) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Beckman Coulter Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-01011-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-10-19
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    A software defect has been identified on newly installed au480 systems with software version 1.81. the defect may generate incorrect test results for 2 step method assays on au480 systems configured with closed channels. the software defect occurs when edit and then confirm are pressed on the specific test parameters screen of closed channels. the result is that the measuring point-2 settings will be deleted. the measuring point 2 settings are used for sample blank adjustments for 2 step methods.For samples which may have lih (lipemia, icterus, hemolysis) or other interferant, there is a potential for false high results to be calculated for 2 step method assays on closed channels.
  • 조치
    A Beckman Coulter FSE will visit each laboratory and change the closed channels to open channels. A set of action steps and associated instructions is also provided in the Customer Letter which Laboratories can use to determine if they have been affected. This action has been closed-out on 22/08/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    AU480 Clinical Chemistry Analyser - software version 1.81(Systems configured with Closed Channels ONLY) Part Numbers: B12183, B11810; B71362ARTG Number: 236636
  • Manufacturer

Manufacturer