AvantGuard 1600 Li160A0(Hospital bed) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Hill-Rom Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00253-1
  • 사례 위험등급
    Class III
  • 사례 시작날짜
    2016-03-08
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Hill-rom has received reports that the backrest actuator on the avantguard 1600 beds may experience a jolted motion, noise (clanking or chattering), premature wear, or backrest functionality loss when a load is applied to the backrest section. these issues do not affect the cpr functionality in case of emergency.The reported issue is unlikely to cause adverse health consequences.
  • 조치
    Hill-Rom has developed a correction and will contact the affected facilities to modify the affected beds. The CPR function is still operational in case of emergency. For this reason, the device can continue to be used whilst awaiting the correction. If however, the unit experiences a loss of backrest functionality, users are advised to take it out of service and contact Hill-Rom.

Device

  • 모델명 / 제조번호(시리얼번호)
    AvantGuard 1600 Li160A0(Hospital bed)Models: CS160A0005XXX & PRS160A0005XXXSerial Numbers: HRP001288830 to HRP002497322Manufactured between June 2009 and December 2015.ARTG number: 127268