AVOXimeter 1000E Cuvettes (used for oximetery assessment during heart catheterisation procedures). An in vitro diagnostic medical device (IVD). 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Zoll Medical Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01265-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-09-29
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    It has been determined that a limited number of avoxmeter 1000e cuvettes have exceeded internal control limits for gap measurement. the gap measurement is incorporated into the pathlength value, which in turn is entered into the settings of avoximeter instruments when a new lot of cuvettes are used to run a test. the pathlength values add an additional degree of accuracy in the total haemoglobin (thb) reported value and those calculations where thb is used as a variable. the pathlength values do not affect %02hb, %cohb or %methb results.The affected avoximeter cuvettes may have a slight upward shift in pathlength that can result in a clinically insignificant corresponding downward shift in thb. in a worst case, this shift may cause the reporting of a thb result approximately 5% lower than the actual thb value (e.G. an actual value of 10 g/dl is reported as 9.5 g/dl). the shift presents a low risk to health in the intended clinical uses associated with the avoximeter system.
  • 조치
    Zoll Medical is advising users of the possibility of a downward shift in tHb values with the affected lots.

Device

  • 모델명 / 제조번호(시리얼번호)
    AVOXimeter 1000E Cuvettes (used for oximetery assessment during heart catheterisation procedures). An in vitro diagnostic medical device (IVD).Product Code: C100BLot Numbers: 607-108, 607-109, 607-110, 607-111, 607-112, 607-114, 607-115, 607-116, 607-117, 607-118, 607-119, 607-120, 607-122, 607-123, 607-124ARTG Number 262423
  • Manufacturer

Manufacturer