AXIOM Artis, Artis zee and Artis Q/Q.zen system (fluoroscopic angiographic x-ray system) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01004-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-07-28
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens has identified that contaminants in the form of biomass can develop in the cooling system of artis systems, which can result in damage to the pump system. the presence of contaminants in the tube cooling circuit could impair the performance of the pump. the tube assembly can become overheated so that no radiation can be released. there is a risk that an ongoing procedure cannot be continued, however, the probability of this occurring is unlikely.
  • 조치
    Siemens is advising users that a technician will be installing an additional external filter in the cooling water circuit to prevent the failure of the pump system. A further correction will be implemented to prevent the development of biomass in the cooling circuit as a permanent correction. Siemens is advising that there is no requirement to re-examine patients.

Device

  • 모델명 / 제조번호(시리얼번호)
    AXIOM Artis, Artis zee and Artis Q/Q.zen system (fluoroscopic angiographic x-ray system)Catalogue/Lot Numbers: 5904433, 5904649, 5904466, 5904441, 5904656, 7728350, 7007755, 7412807, 7727717, 7728392, 7413078, 5917054, 7555373, 7555357, 7008605, 7555365, 10502501, 10848280, 10848353, 10502502, 10848281, 10848354, 10502505, 10848283, 10502504, 10848282, 10848355, 10502507, 10094135, 10094142, 10094137, 10094143, 10094141, 10094139, 10280959ARTG Numbers: 102173, 27395, 102177