Babytherm 8004/8010 and Babytherm 8000WB for newborns (Infant warmers and open infant care units) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Draeger Medical Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00961-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2014-09-09
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Dräger have become aware of a case in which a side panel on a babytherm 8010 did not fully slide to the bottom in the corner guide and was not completely locked. the side panel did not open during this incident and no patient was affected. as part of the review, dräger identified that the inner 70mm panels were not always used however, the inner 70mm panels are part of the safety concept of the babytherm 80xx. if contrary to the information provided in the instructions for use manual the inner panels are not inserted and if a side panel has not been fully locked into position this may allow the baby’s movements to open the side panel. it may be possible for a child to fall out.
  • 조치
    Users are reminded to pay attention to the information provided in the Instruction for Use Manual: i.e., “Check that the side panels lock securely into position and that they remain securely locked”. If the locking mechanism is not functioning or the inner panels are missing or damaged, please contact Dräger Customer Service to arrange correction. This action has been closed-out on03/02/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    Babytherm 8004/8010 and Babytherm 8000WB for newborns (Infant warmers and open infant care units)ARTG Number: 121161
  • Manufacturer

Manufacturer