balanSys Knee System (CR, UC, PS and RP) balanSys Bone Orientated 4in1 SMaRT Instruments, balanSys BICON Knee System Template, balanSys PS Knee System Template, balanSys PE Trial Inlay 67/8-20.5, balanSys PE Trial Inlay 67/8-23, balanSys CR/PS Tibial Template 67, balanSys RP Trial Tibial Plateau 67 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Mathys Orthopaedics Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00094-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-01-31
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The incorrect size determination table in surgical technique indicates that the femoral component size c is compatible with a tibial inlay/tibial plateau size 67. however, tibial components size 67 may be correctly used only with the smaller femoral components sizes a or b. the incompatible combination of the balansys cr, uc and ps systems (fixed bearing)may result in the following risks: - there is an immediate risk that the knee cannot be optimally equilibrated, which may lead to pain and instability. - in the long term, there is a risk of increased wear of the polyethylene, which may lead to pain, osteolysis, tibial loosening, instability and revision surgery. in the balansys rp system (rotating platform), there is a risk of increased soft-tissue irritation associated with pain.
  • 조치
    Mathys Orthopaedics is advising surgeons and hospitals that the Surgical Technique is incorrect as it indicates that the femoral conmponent size C is compatible with a tibial inlay/tibial plateau size 67. Tibial components size 67 should only be used with smaller femoral components size A and B. A single patient was implanted with the incorrect combination in Australia. The associated surgeon has been contacted and provided with additional information for the ongoing management of the patient.

Device

  • 모델명 / 제조번호(시리얼번호)
    balanSys Knee System (CR, UC, PS and RP)balanSys Bone Orientated 4in1 SMaRT InstrumentsArticle No: 336.030.053balanSys BICON Knee System TemplateArticle # 330.030.034balanSys PS Knee System TemplateArticle # 330.030.035balanSys PE Trial Inlay 67/8-20.5 Article # 71.34.0222 - 71.34.0227balanSys PE Trial Inlay 67/8-23Article # 71.34.0263 - 71.34.0269balanSys CR/PS Tibial Template 67Article # 71.34.0538balanSys RP Trial Tibial Plateau 67Article # 71.34.0296
  • Manufacturer

Manufacturer