Bard Arctic Sun 5000 Temperature Management System 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Bard Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-01116-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-11-17
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Bard medical division (bmd) has identified that the arctic sun 5000 temperature management system with certain serial numbers may contain electronic components that lead to a premature drainage of the internal control panel coin cell battery. the exhausted battery will prevent the arctic sun temperature management system control panel from launching software and initiating operation. this presents a potentially serious issue since the system is used to control patient temperatures within a pre-set range and has the potential of not being available to provide therapy when needed.The issue identified however does not affect an arctic sun 5000 once it is successfully powered up since the batteries are no long being called upon and the unit is operating on mains power.
  • 조치
    The sponsor is advising the user that if the device is currently in clinical use, it can continue to be used without concern for this failure. The battery drainage issue only occurs when the unit is off and in stand-by mode. The sponsor will arrange for a technician to come to the user's facility for the evaluation and potential replacement of the affected electronic components. This action has been closed-out on 26/08/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Bard Arctic Sun 5000 Temperature Management System Serial numbers: 1255, 2325, 2515, 2530, 2643, 2650, 2722, 2755, 2780, 2781, 5157, 5178, 5181, 5240, 5374, 5459, 5461, 5232. ARTG Number: 150422
  • Manufacturer

Manufacturer