Baxter Irrigation solutions 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Baxter Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00329-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-04-10
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    An investigation into an adverse event report concerning a 6l renal dialysis solution that affected a single patient revealed the solution had been diluted. baxter healthcare is recalling the irrigation solutions listed above as further investigations revealed that these products were also potentially diluted with water for injection.
  • 조치
    Customers are advised to locate and remove affected product from potential use. The TGA has published a web statement regarding this recall which is available at https://www.tga.gov.au/alert/dianeal-peritoneal-dialysis-solution-and-various-irrigation-solutions

Device

  • 모델명 / 제조번호(시리얼번호)
    Baxter Irrigation solutions1.5% Glycine IrrigationCode: AHB7316Batch: S63S1Expiry: 30/09/20150.9% Sodium Chloride for Irrigation Code: AHB7616Batch: S66P2Expiry: 30/11/2015 0.9% Sodium Chloride for IrrigationCode: AHB7616Batch: S71N8Expiry: 31/03/20161.5% Glycine Irrigation SolutionCode: AHB7317Batch: S74P6Expiry: 31/05/20160.9% Sodium Chloride for Irrigation Code: AHB7127Batch: S82F6Expiry: 31/10/2016ARTG number: 155928
  • Manufacturer

Manufacturer