BCS / BCS XP Automated Blood Coagulation Analyser with Dade Innovin .An in vitro diagnostic device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00025-2
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-01-08
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens healthcare diagnostics has observed that in very rare cases false short clotting times for pt on bcs and bcs xp with dade innovin can occur. these false short clotting times were observed with turbid, haemolytic and/or icteric samples from intensive care patients. the corresponding reaction curves of these rare samples showed a pre-peak, which can lead to falsely short clotting time. in rare cases the result was not flagged by the existing check algorithms.There is a potential to miss a prolonged prothrombin time which could lead to a delay of intervention and a risk of a significant bleed.
  • 조치
    Siemens is advising users check the inference levels given in the respective Application Sheets of the BCS and BCS XP systems, as such an effect can occur in turbid, haemolytic or icteric samples from intensive care patients. Siemens will provide a software upgrade which can identify irregular reaction curves in the near future. This action has been closed-out on 27/01/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    BCS / BCS XP Automated Blood Coagulation Analyser with Dade Innovin .An in vitro diagnostic device (IVD)BCSSiemens Material Number (SMN):10454728, 10454729, 10454742, 10459303,10460659, 10461881 BCS XPSMN: 10459330, 10462449, 10461894, 10470625 Dade InnovinSMN: 10284500, 10445704, 10445705, 10445706, 10465673, 10465674ARTG Numbers: 178116 & 178501