Bending/Cutting Pliers with Nose, for LOCK Plates 2.0 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Synthes Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-01359-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-12-23
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The pliers have the potential for tip breakage while bending the extra-large compact 2.0 mandible locking plates (part numbers 447.412, 447.420, 447.422, 447.423, 447.424, 447.425, 447.426, 447.412s, 447.420s, 447.422s, 447.423s 447.424 s, 447.425 s and 447.426s ) with the last hole feature of the instrument.If a nonconforming bending/cutting pliers with nose for lock plates 2.0 is used and breaks while in the operating room, surgical delay may occur while alternate bending or cutting instruments are located.To date, there have been no reports of adverse events related to this issue. such reports are unlikely to occur as the bending/cutting pliers with nose for lock plates 2.0 are not used in direct contact with the patient but rather on the plates before they are inserted.
  • 조치
    Customers are requested to:- 1. Check inventory immediately for the affected units; 2. Synthes will contact customers to arrange a replacement or credit as available; 3. An interim inspection procedure had been provided to customers whilst awaiting replacement stock. This action has been closed-out on 14/07/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Bending/Cutting Pliers with Nose, for LOCK Plates 2.0 Part Number: 329.143 Multiple Lot Numbers affectedARTG Number: 153950