Biograph mCT and Biograph mCT Flow systems running VG50A software and equipped with the Adaptive 3D Intervention option 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-01243-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2014-11-21
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens has identified a potential malfunction that may cause a display freeze during ct interventional procedures. the user interface may lock-up when performing an adaptive 3d interventional procedure when the "auto save at last displayed images as key images" is enabled. when the issue occurs, normal interaction with the system is no longer possible and no error messages are displayed. therefore the end-user and/or interventionalist may not know that the system has stopped interacting and continue wiht the interventional procedure.
  • 조치
    Siemens is advising their customers to reconfigure the system Storage to deactivate "Auto save at last displayed images as key images". A software update will be implemented to correct the problem. This action has been closed-out on 18/03/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Biograph mCT and Biograph mCT Flow systems running VG50A software and equipped with the Adaptive 3D Intervention optionSoftware version: VG50AARTG number: 144218
  • Manufacturer

Manufacturer