Biograph, Symbia T and Symbia Intevo SPECT/CT systems 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00654-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2018-06-05
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens healthineers are aware of potentially incorrect tube-current calculations by the care dose4d algorithm when utilising only pa or ap topograms. depending on the geometrical shape of the skull, rare cases may occur in which the calculated dose distribution is not appropriate for scans that include the head, when based only on pa or ap topograms. the care dose4d software may select the maximum tube current for the uppermost part of the skull, which incidentally leads to unnecessary radiation exposure.
  • 조치
    Siemens is advising that the manufacturer is developing appropriate solutions to address the problem with high priority. The described issue does not occur when using a lateral topogram. Accordingly, Siemens strongly recommend the utilization of topograms in the lateral position for all CT and PET/CT scans that include the entire head. Alternatively, if users choose to perform both a lateral topogram, in conjunction with a PA or AP topogram they must ensure that they perform the lateral topogram last as CARE Dose4D is based on the last topogram. If users do not utilize a lateral topogram for scans including the entire head, Siemens strongly recommend that users deactivate the CARE Dose4D feature.

Device

  • 모델명 / 제조번호(시리얼번호)
    Biograph, Symbia T and Symbia Intevo SPECT/CT systemsCatalogue Numbers: 8728581, 10097303, 10097289, 10097302, 10532746, 10532748, 10528958, 10529158, 10529160, 10529161, 10528955, 10507786, 10507786, 10534160, 10248669, 10248672, 10248673, 10248670, 10525581, 10764804, 10764802, 10764803, 11007962, 10764801, 10275007, 10275010, 10275008, 10275009 and 8717733.ARTG Numbers: 123883, 144218 and 271560.