Biosense Webster PENTARAY NAV and PENTARAY NAV eco Catheters (used for electrophysiological mapping of the heart) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Johnson & Johnson Medical Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00488-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-04-20
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The manufacturer, biosense webster, is clarifying the contraindication language in the instructions for use (ifu) and product labelling for use of this catheter relative to patients with prosthetic valves. the current language in the ifu provides a precaution against use of the pentaray catheter in patients with prosthetic valves under the contraindication section stating: “[the] use of this catheter may not be appropriate for use in patients with prosthetic valves.” the update to the ifu is to clarify the contraindication statement to the following: “do not use pentaray catheters in patients with prosthetic valves”.If excessive force is applied on the entangled catheter spine, there is a potential for parts to detach and embolise inside the patient’s body, which may lead to serious complications like stroke, transient ischemic attack, myocardial infarction or pulmonary embolism. the likelihood of these serious complications remains low.
  • 조치
    Johnson and Johnson Medical (JJM) are providing users with an updated IFU to follow to ensure the products are not used in patients with prosthetic valves. This action has been closed-out on 10/05/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    Biosense Webster PENTARAY NAV and PENTARAY NAV eco Catheters (used for electrophysiological mapping of the heart)Catalogue Numbers: D128201, D128202, D128203, D128204, D128205, D128206, D128207, D128208, D128209, D128210, D128211, D128212All Lot Numbers affectedARTG numbers: 129479, 203362, 231772
  • Manufacturer

Manufacturer