BiPAP A 40 Ventilatory Support System; Detachable Battery Module, USA/INTL 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-01438-1
  • 사례 위험등급
    Class III
  • 사례 시작날짜
    2017-11-27
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Philips respironics has identified an issue where a non-conforming part (side panel) was utilized on the bipap a40 that affects the mechanical connection between the bipap a40 and the detachable battery module (part number 109677). the issue results in the inability to connect the bipap a40 device to the detachable battery module during setup before putting the ventilator into operation on a user; therefore, there is no safety risk to the patient. the impacted bipap a40 units continue to work properly in configurations without the detachable battery module. a non-conforming part was utilized on the bipap a40 that affects the mechanical connection between the bipap a40 and the detachable battery module.
  • 조치
    1. Complete the inspection of the devices for the potentially faulty side panel per the inspection instructions provided with the letter. 2. Complete and return Business Reply Form (BRF) and list the total quantity inspected, the quantity of correct devices, and the quantity of incorrect devices. 3. Contact your local Customer Service who will provide direction on how to return your device(s) for any confirmed faulty side panel replacement.

Device

  • 모델명 / 제조번호(시리얼번호)
    BiPAP A 40 Ventilatory Support SystemProduct code: 1111171Detachable Battery Module, USA/INTLProduct code: 1096770ARTG number: 200289