BK Virus R-gene kit when used on SmartCycler 2.0 platform. An in vitro diagnostic medical device (IVD). 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Biomerieux Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00141-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-01-25
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Following a customer complaint and an internal investigation, biomérieux has confirmed that when testing weak viral load samples, in conjunction with bk virus r-gene kit on smartcycler 2.0 platform, negative results may be obtained instead of positive as expected. the investigation confirmed that the issue has been observed only with smartcycler 2.0 and the issue only concerns low viral load samples (viral load equal or below the sensitivity control). the issue is detected by the sensitivity control (sc) that is the equivalent of a weak positive sample, which will be systematically negative instead of positive in most of the cases. when no sensitivity control is systematically run, there is a potential risk to obtain a false negative result. based on the investigation data, the performances claimed in the ifu regarding the use of smartcycler 2.0 in conjunction with bk virus r-gene kit could not be warranted.
  • 조치
    bioMérieux is advising customers to stop using the BK Virus R-gene kit ref. 69-013B on Smart Cycler 2.0. It is recommended that any concern customers have regarding previously reported results is discussed with the Laboratory Medical Director to determine the appropriate course of action.

Device

  • 모델명 / 제조번호(시리얼번호)
    BK Virus R-gene kit when used on SmartCycler 2.0 platform. An in vitro diagnostic medical device (IVD).(used for measuring the viral load of BK virus in whole blood, plasma and urine samples)Catalogue Number: 69-013BARTG Number: 211361
  • Manufacturer

Manufacturer