Brain and Spinal Cord Stimulators 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Medtronic Australasia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00289-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2014-03-12
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Cycling is a programmable feature which allows the stimulation output of a device to toggle on and off based on programmed settings. enabling cycling at certain parameter settings may decrease the recharge interval for rechargeable devices and the device longevity.In may 2012, the first two cases of premature battery depletion where cycling was programmed at fast rates (0.1s on/off) in conjunction with low patient amplitudes (<3 volts) were reported. both of these reports involved the primeadvanced device model 37702. both devices were explanted, returned and analysed. this inspection, engineering bench testing, and detailed understanding of how the patients used these devices confirmed that specific patient therapy parameters in conjunction with cycling being enabled resulted in a higher current drain on the battery when compared to running the device in continuous mode. higher current drain results in reduced battery longevity.
  • 조치
    Surgeons are advised that the updated calculations and graphs available at professional.medtronic.com/cycling to assess the affect of cycling on device longevity. Surgeons are requested to remind their patients to continue to check battery status. For more details, please see http://www.tga.gov.au/safety/alerts-device-medtronic-neurostimulation-140317.htm .

Device

  • 모델명 / 제조번호(시리얼번호)
    Brain and Spinal Cord StimulatorsMultiple model numbers affectedARTG Numbers: 160118, 160117, 188274, 188034, 141861, 154866, 177594, 141860, 215748, 215749, 215750 and 215751
  • 의료기기 분류등급
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.