Brainlab Cranial Navigation System (all versions before Cranial 3.0) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Brainlab Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00558-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2015-06-23
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Brainlab has detected that, when using the brainlab cranial navigation system, the following may have a significant effect on the overall navigation accuracy: • large distance between reference array and region of interest • major changes of the camera position relative to the reference array during the procedure those instances could potentially intensify small inaccuracies arising from individual steps of the complex navigation procedure. in the worst case scenario, these inaccuracies may cause an inaccurate display of instruments by the navigation system, compared to the actual patient anatomy. if these inaccuracies are not detected by user verification of navigation accuracy as described in the user manual, this could ultimately lead to ineffective treatment, serious injury or even death of the patient.
  • 조치
    Brainlab has developed a software update to improve navigation accuracy. Brainlab will contact their customers to arrange for the installation of the update. This action is a follow-up to RC-2013-RN-00397-1. This action has been closed-out on 05/12/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Brainlab Cranial Navigation System (all versions before Cranial 3.0)Product Number: Brainlab Cranial Navigation System (all versions before Cranial 3.0)ARTG Number: 96517