Brainlab Offset Cup Impactor Universal (blue handle) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Brainlab Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00178-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-02-07
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    An issue has been detected that when the brainlab offset cup impactor universal (blue handle) is reprocessed in a sterilisation tray as described in the current brainlab cleaning, disinfection and sterilisation guide (revision 5.2 and previous versions), the device may not be determined as sterile. the required sterility assurance level (sal) of 10-6 can only be achieved if the device is individually wrapped for sterilisation. if a non-sterile brainlab offset cup impactor universal (blue handle) is used during a surgical procedure to implant an acetabular cup, contaminated residues on the product surface could, directly or indirectly, enter the patient’s body, potentially leading to an infection. at this stage, there have been no reports of affected patients due to this issue.
  • 조치
    Brainlab is advising users to follow the sterilisation instructions in the current version (BL-IL-60960-69ML Rev. 1) of the IFU. Users are reminded that in addition to following the other instructions in the IFU, the Offset Cup Impactor Universal (blue handle) must be sterilized in an individually wrapped configuration.

Device

  • 모델명 / 제조번호(시리얼번호)
    Brainlab Offset Cup Impactor Universal (blue handle)Catalogue Numbers: 52856, 52856AARTG Number: 121094