Brianlab iPlan RT Image (versions below v.4.1.2) / iPlan RT and BrainSCAN (discontinued in 2002) (Radiation therapy treatment planning system) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Brainlab Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-01021-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2015-10-20
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The only image sets that are supported by the brainlab radiation treatment planning software even when containing varying calibrations of pixel values in between image slices are ct image sets & pet image sets normalised to suv with suv also loaded in and displayed by the brainlab swif such an unsupported image set is used the same pixel value calibration as contained in the middle image slice of the set to all slices is applied. in a pet image set uniform calibrated pixels might then have a different activity value, thus displayed with a different brightness and colour, compared to the image slices as displayed by the scanner using varying calibrations. if this effect occurs when using an unsupported image set, it can be misleading to the user during treatment planning and result in an incorrect definition of vols of interest in the planning s/w. the objects created for these vols of interest might differ in location or dimension from the actual desired definition of the structure.
  • 조치
    Brainlab will provide a software solution with the corresponding new feature for affected iPlan RT / iPlan RT Image customers to prevent the described scenario from occurring. In the interim, customers are advised to: - Except for CT and PET used with SUV, only load and use image sets in iPlan RT / iPlan RT Imagethat have the same pixel value calibration for all images of the set. - Before using image datasets loaded into the Brainlab Radiation Treatment Planning Software, always compare the images displayed in the Brainlab application to the images as displayed by the imaging device (scanner). Delete and do not use the image sets within iPlan RT in case any deviation is detected and contact Brainlab support. Brainlab will not provide a software update for BrainSCAN. Brainlab strongly recommends customers still using BrainSCAN clinically to phase out the use of this software as soon as possible but in no event later than January 2016. This action has been closed-out on 02/12/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Brianlab iPlan RT Image (versions below v.4.1.2) / iPlan RT and BrainSCAN (discontinued in 2002) (Radiation therapy treatment planning system)ARTG Number: 121114