BrightView X/XCT/XCT Upgrade 882480: BrightView (version 1.2.3 or 1.2.4)882478: BrightView X (version 2.5.3 or 2.5.4)882482: BrightView XCT (version 2.5.3 or 2.5.4)882454: BrightView XCT Upgrade (version 2.5.3 or 2.5.4) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00051-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-01-18
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Philips have identified 4 issues with the brightview systems:1. during a cardiac non-auto body contouring (abc) scan, using cardiac high resolution (chr) collimators, the collimator may come in contact with the bottom edge of the patient pallet. if this occurs, collision sensors would pause the scan. although unlikely, if the collision sensors malfunction & the operator fails to activate the e-stop, this could lead to serious bodily injury for the patient.2. during a scan, the gantry motion may halt prematurely. this causes the scan sequence to pause.3. operator will receive a warning message that the scan may not complete & re-mark the points used to define the orbit when defined with a detector relative angle of 90 & head first supine/feet first prone setting. 4. resolved defects section of user release document 1.2.4/2.5.4 for brightview, brightview x & xct listed 2 pre-programmed motions (ppms) whose motion sequences were changed by the release of s/w v1.2.4/ 2.5.4.
  • 조치
    The sponsor will provide a system software upgrade to correct the reported issues.