BrightView XCT and BrightView X- upgraded with the XCT Flat Panel Detector (Gamma camera) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00682-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-06-27
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    During a daily xct quality assurance phantom scan for image quality and hounsfield unit (hu) linearity an operator experienced resistance when engaging the locking handle of the fpd in the deployed position. with the locking handle in the locked position, the flat panel detector (fpd) locking mechanism was not locked, moved unexpectedly and contacted the imaging detector during system movement. philips’ investigation determined that the internal linkage shaft broke preventing the fpd from being locked into its correct deployed position.
  • 조치
    Philips Healthcare Field Service Engineers will undertake a field corrective action to ensure no unexpected movement of the Flat Panel Detector. Philips is advising their customers to cease use of the XCT Flat Panel Detector until the implementation of the field corrective action. This action has been closed-out on 18/02/2016.