Brilliance 64 system with software version 4.00xx516 and Ingenuity CT system running software version 4.00xx518 (full body CT systems) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2012-RN-01187-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2012-11-23
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    1. during an interventional procedure with an interventional exam card reconstruction increment set to a negative value, one exposure of cct was not reconstructed – there was radiation but images were not updated on the screen. 2. when moving the couch using the gantry buttons during an interventional procedure & the user chooses an interventional exam card with the reconstruction increment set to a negative value, the user has to wait a few seconds for the new position to update. there is an image update delay at about 5 second intervals. 3. when reconstructing images with o-mar the system may fail to provide a non-omar set of images as required. the non-omar images must be manually reconstructed. 4. the rotor begins to spin without waiting for service personnel to request the go command. 5. incorrect cbv values in the brain perfusion application may occur if the user takes brain perfusion images from a ge scanner into a specific application installed on a philips scanner.
  • 조치
    Philips is provided work around instructions for the issue and will be implementing a software update to correct the problem.