Brilliance Big Bore Oncology CT and Radiology CT (with software version 4.2.0)(4D CT pulmonary phase issue) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00370-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-03-16
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    An issue has been identified where a retrospective respiratory gated 4d ct scan is performed with the bellows pulmonary gating device, the ct images might be reconstructed at a single phase, while the annotations on the images incorrectly indicate that the reconstruction is from the requested specific phases (0% to 90%). this may lead to a couple of scenarios that may result in misrepresentation of the 4d ct images with a respiratory gated 4d ct scan and reconstruction. in the event this occurs a misrepresentation of ct images may result, due to the incorrect image pulmonary phase annotation or the need to rescan the patient due to an incorrect image pulmonary phase.
  • 조치
    Philips is recommending users to follow the instructions as outlined in the recall notification letter. The letter contains information to prevent the occurrence of the scenarios as well as the corrective actions to take when they do occur.

Device

  • 모델명 / 제조번호(시리얼번호)
    Brilliance Big Bore Oncology CT and Radiology CT (with software version 4.2.0)(4D CT pulmonary phase issue)ARTG Number: 98868
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • Source
    DHTGA