Brilliance Big Bore Oncology CT, Brilliance Big Bore Radiology CT with software version 4.2.0Used with associated Pinnacle3 Tumor LOC server (version 14.0) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00085-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2018-01-30
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Philips became aware of issues with brilliance big bore ct systems running software version 4.2.0. the most significant issues are:1- efov (extended field of view) check message does not display when fov is set >600mm for a paused scan2- prefer “center x/y 0” and “disable result rotation” checkboxes checked automatically when checking the "ct simulation"3- tumor loc allows poi (point of interest) locking when a non-isocenter slice is displayed4- tracker scan completed, no images were generated5- continuous link plans the clinical scan outside the surview6- applying tilt on first mpr result, when the other axial results are unplanned causes change in scan length and number of cycles of scan7- “system is still initializing. please wait…” error8- auto locations not functioning as expected9- system ended exam before completion10- system freeze-up after pause is pressed.
  • 조치
    Philips is releasing a Field Change Order (FCO) to correct these issues. A Philips Field Service Engineer will contact users to schedule the FCO updates. In the interim, users are advised to refer to the Customer Letter for respective actions in regards to the outlined issues.