Brilliance CT Big Bore Oncology, Brilliance CT Big Bore Radiology and Brilliance CT 16 (Air), software version 3.6.0 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-00941-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2013-09-06
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    There are artifacts appearing on the tracker images during threshold-triggered bolus scans with protocols using a rotation time of 0.4 seconds. in those cases only, the scan might be triggered either too soon or too late, which may lead to inappropriate contrast opacification during the clinical scan. if the acquired images are not diagnostically acceptable, a rescan of the patient need to be performed.
  • 조치
    Clinical judgment should be used by the customer/user to determine if the images acquired in the dataset contain enough information to make a diagnosis. Philips Healthcare is implementing software update version 3.6.2 to correct the issues.