Brilliance CT series (6, 10, 16, 16 Power, 40, 64, Big Bore, iCT, iCT SP), Ingenuity CT, Ingenuity Core, Ingenuity Core128, Ingenuity Flex, MX8000 Dual v. EXP 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00475-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2014-04-24
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Philips have found that if the service latch is not properly secured during servicing of the device, the table top's subframe becomes free floating causing unintended horizontal motion during use. there is a risk of serious injury to the patient and operator/technician as a result of the table top free floating. for the patient: - there is a potential for disconnection or movement of invasive medical devices such as i.V.’s, tracheostomies, and surgical drains for the operator/technician: - there is a potential for pinch point on the subframe; and - a potential for entrapment between the table and the gantry if the operator/technician stands between them to shift the patient forward.
  • 조치
    Philips is advising their customers to contact their Field Service Engineer, if horizontal free-floating couch motion is experienced. This action has been closed-out on 11/04/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Brilliance CT series (6, 10, 16, 16 Power, 40, 64, Big Bore, iCT, iCT SP), Ingenuity CT, Ingenuity Core, Ingenuity Core128, Ingenuity Flex, MX8000 Dual v. EXPARTG Number: 98868
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • Source
    DHTGA