Brilliance iCT and Brilliance iCT SP with software versions 4.1.0.xxxxx, or 4.1.1.xxxxx, or 4.1.2.xxxxx 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00068-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-01-28
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Philips has identified a software defect that may result in the scanner updating the planned scan length to be longer than the user intended. this may happen if:1. the exam is planned with low pitch and rotation time values, for example, 0.3 pitch and 0.27 rotation time, and2. the plan scan length is increased or lengthened on the surview/scout, and3. the system displays a pop-up message that the current pitch, rotation time, and planned scan length are incompatible and suggests a new pitch and rotation time. 4. if the user accepts the suggested pitch and rotation time changes, the system adjusts the plan to the maximum allowed scan length, which will plan the study beyond the original intended length.
  • 조치
    Philips is providing instruction to assist identify if a system is affected and advising end users to ensure that the scan length is verified prior to initiating the scan. A Philips Field Service Engineer will contact affected customers to arrange for the software to be updated. This action has been closed-out on 19/07/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Brilliance iCT and Brilliance iCT SP with software versions 4.1.0.xxxxx, or 4.1.1.xxxxx, or 4.1.2.xxxxxARTG Number: 98868
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • Source
    DHTGA