Brilliance iCT and iCT SP up to and including Software Version 3.2.5 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-00766-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2013-07-24
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    When using 80kvp with intravenous contrast in head and neck studies with u-filters (ua, ub, uc), a vascular artefact that resembles thrombus may appear on the image. there is a risk of misdiagnosis which can lead to unnecessary treatment of a patient. for certain patient populations, this could contribute to serious injury.
  • 조치
    When the settings (80 kVp with intravenous contrast in head and neck studies with U-filters (UA, UA and UC)) are used, Philips recommends performing additional reconstruction with a non-U filter and comparing the result images to verify no such artefact is evident. Customers are advised to continue to use 80 kVp with U-filters for brain perfusion scans. Refer to the Instructions for Use (IFU) for the selection of an arterial input unaffected by discontinuity should one occur.

Device

  • 모델명 / 제조번호(시리얼번호)
    Brilliance iCT and iCT SP up to and including Software Version 3.2.5ARTG Number: 98868
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • Source
    DHTGA