BVM (Bag-Valve-Mask) Manual Resuscitation Systems 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Intersurgical Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00058-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2018-01-16
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    A manufacturing fault has been identified with the intersurgical bvm manual resuscitation system, which could result in the directional valves positioned on the back plate of the self inflating bag being orientated incorrectly. the incorrect orientation of the directional valves could result in the bvm manual system becoming over pressurised, which could result in patient barotrauma.
  • 조치
    Intersurgical is advising users to inspect stock and quarantine affected units for return. In instances where an alternative product is not available, Intersurgical is advising users to inspect each device before use to determine if it is affected by the issue. If the device is not affected, it can be safely used.

Device

  • 모델명 / 제조번호(시리얼번호)
    BVM (Bag-Valve-Mask) Manual Resuscitation SystemsReference Number: 7150000, 7151000, 7152000, 7153000, 715400Lot Numbers: 7150000 – 372894, 7151000 – 372886, 7152000 – 372893, 7153000 – 372892, 7154000 – 370753 ARTG Number: 225855 (Intersurgical Australia Pty Ltd - Resuscitator, pulmonary, manual, single use)