CADD- Solis Ambulatory Infusion Pumps 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Smiths Medical Australasia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00447-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-05-28
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Smiths medical has become aware of an issue with an intermittent occurrence of binding of the locking assembly on some cadd-solis pumps. binding of the cassette/ keypad lock can occur after latching the cadd-solis medication cassette reservoir or administration set “disposable” to the pump. when binding occurs, it can prevent the key from fully rotating the cassette/ keypad lock to the locked (engaged) position. to start an infusion with cadd-solis pump models 2111(grey keypad) and 2112 (yellow keypad), and model 2120 (vip), when in pca mode, the lock must be fully engaged. cadd-solis pump model 2120 will start in all other delivery modes without the cassette/ keypad lock being engaged.Smiths medical has received no reports of serious injury or death related to this issue.
  • 조치
    Smiths Medical has resolved the problem associated with the binding of Cassette/ Keypad Locks and implemented a lubrication step during manufacture and during servicing at Smiths Medical Authorised Service Centres. A customer service representative will contact users to arrange for repair of their CADD-Solis Pumps. This action has been closed-out on 08/08/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    CADD- Solis Ambulatory Infusion Pumps Model Numbers 2111, 2112, and 2120Serial Numbers: 1061043 through 1067598 ARTG Number: 109467
  • Manufacturer

Manufacturer