Calibrator E used for calibration of the ADVIA Centaur Systems Cortisol, Progesterone and Testosterone assays 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Ltd Diagnostics Division 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-01060-1
  • 사례 위험등급
    Class III
  • 사례 시작날짜
    2014-10-07
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens healthcare diagnostics has identified a positive bias for calibrator e kit lots ending in 38 and 39 with the advia centaur systems cortisol and progesterone assays compared to the respective master curves. patient serum samples evaluated with the advia centaur cortisol assay and calibrated with affected lots of calibrator e kit demonstrate an average positive bias of 16% across the assay range with a maximum bias of 28%. patient serum samples evaluated with the advia centaur systems progesterone assay and calibrated with affected lots of calibrator e kit demonstrate a positive bias with samples above 30 ng/ml. the average positive bias on the advia centaur/advia centaur xp is 20% with a maximum bias of 39%. the average positive bias on the advia centaur cp is 17% with a maximum bias of 35%.The advia centaur systems testosterone assay is not affected by this correction.
  • 조치
    Siemens is notifying their customers of the potential for positive bias when using the affected calibrators. As the bias only occurs outside of the useful clinical range the calibrators can continue to be used. Unaffected calibrators can be ordered if required. This action has been closed-out on 01/03/2016.