CARDIOSAVE Hybrid IABP Intra-Aortic Balloon PumpCARDIOSAVE Hybrid CARDIOSAVE Hospital Cart CARDIOSAVE Hospital Cart RoHS 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Getinge Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00603-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2018-06-29
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Getinge has received complaints involving the cardiosave iabps regarding ingress of fluids into the iabp affecting various electronic circuit boards. this situation would prevent initiation or continuation of therapy. to date, getinge has received one report of an adverse event in which one death was associated with a saline spill / liquid ingress.
  • 조치
    Getinge are advising that affected sites will be contacted to schedule an on-site service of the Cardiosave IABP to install a protective top cover and perform a System Functional Test. Until Getinge have implemented the corrective actions above, users should adhere to the instructions outlined on the Customer Letter when using the Cardiosave IABP

Device

  • 모델명 / 제조번호(시리얼번호)
    CARDIOSAVE Hybrid IABP Intra-Aortic Balloon PumpCARDIOSAVE Hybrid CARDIOSAVE Hospital Cart CARDIOSAVE Hospital Cart RoHS Part Numbers: 0997-00-0562 and 0997-001-1179ARTG Number: 118266(Getinge Australia Pty Ltd - Circulatory assist unit, pump, balloon)
  • 의료기기 분류등급
  • Manufacturer

Manufacturer