Carescape Monitor B650 (Multi parameter patient monitor) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 GE Healthcare Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2012-RN-00806-3
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2012-08-08
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    1) spo2 saturation limit alarm does not always activate when nellcor spo2 is used in pdm and saturation seconds is not switched on.2) when the carescape monitor b650 is started the monitor may enter into a constant boot loop causing monitor to reboot continuously.3) when the carescape monitor b650 is connected to the unity id and information from a 3-lead ecg cable is transferred through dri (s/5) network, the monitor may reboot.4) upgrading the software for carescape monitor b650 may not update the software for power management card (pmc).5) when the usb cable connected to the carescape monitor b650 is insecurely connected or faulty and is able to move the monitor may reboot. 6) carescape monitor b650 remote printing issue with currently printing prn 50 recorder.7) carescape monitor b650 reboots due to an x server freeze. it may reboot when the monitor is viewed from carescape cic pro or when the printing is initiated from carescape cic pro.
  • 조치
    GE Healthcare is providing users with work around instructions to mitigate the issue. A software and hardware update will be implemented to permanently fix the issue when available.

Device

  • 모델명 / 제조번호(시리얼번호)
    Carescape Monitor B650 (Multi parameter patient monitor)ARTG Number: 115304
  • Manufacturer

Manufacturer