CARESCAPE Monitor B850 (with software versions 1.0.12 and earlier) and CARESCAPE Monitor B650 (with software version 1.1.12 and earlier) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 GE Healthcare Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-01109-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2013-10-31
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Non-invasive blood pressure (nibp) - auto-cycling not starting/stopping, nibp measurement not performed if network time update occurs at same time, infants limits are incorrect. ecg- wrong ecg bandwidth in e-psm(p) or e-(p)re(s)tn module. changing primary ecg lead selection will not change primary measurement in pdm or tram.Central monitor- a temporarily hidden alarm message will not reactivate if different alarm is acknowledged, audible alarm missing after broken network connection, ecg not shown on icentral, standby doesn't end, & incorrect labelling on v leads.Ventilation alarms- incorrect display of airway pressure in bed to bed view, spirometry alarm limits can't be changed.General- alarm off issues in small parameter window, auto-discharge doesn't erase data, invasive pressure lables incorrect.Tram -missing service alarm bed to bed view doesn't show highest st value in s/5 network, missing remote alarm & avoa alarm, monitor disconnect alarms not restored.
  • 조치
    GE Healthcare is advising users of work around instructions to follow as an interim measure. GE is providing users with a laminated sheet summarising the work around instructions to keep near the affected units for easy reference until the software update is implemented to correct the issues. This action has been closed-out on 12/02/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    CARESCAPE Monitor B850 (with software versions 1.0.12 and earlier) and CARESCAPE Monitor B650 (with software version 1.1.12 and earlier)ARTG Number: 115304