Catheters with Beacon Tip Technology 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 William A Cook Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00483-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2016-04-22
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Cook medical has identified an increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. the preliminary investigation has identified that environmental conditions, such as storage temperature, humidity, and the use of vaporised hydrogen peroxide (vhp) for whole-room decontamination within healthcare facilities, may be contributing to the occurrence. it is recognised that there may be other undetermined contributors to this issue and continue to investigate. potential adverse events that may occur as a result of catheter polymer degradation include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. fragments within the vascular system could result in embolisation to the heart or lungs, or occluding blood flow to end organs.
  • 조치
    Cook Medical is advising users to quarantine and return any unused product.

Device

  • 모델명 / 제조번호(시리얼번호)
    Catheters with Beacon Tip TechnologyMultiple products/catalogue numbers affectedAll lots affectedARTG Numbers: 137861, 154909, 153374, 153375,154911, 137861, 214869, 137911, 212573,133996, 208185, 207779
  • Manufacturer

Manufacturer