cCrea A+B Membranes used with ABL800 series blood gas analyser rangeItem # 942-073, 08088 Crea A + 08089 Crea B membran 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Radiometer Pacific Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00537-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-06-24
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    In november 2014, radiometer pacific notified customers (tga ref: rc-2014-rn-01201-i) that the ccrea results for qc decrease to below the lower acceptance limit within the lifetime of the crea membranes. the corrective action implemented at that time was to perform one quality control measurement every 24 hours using s7835 autocheck 6+, level 1 (944-094) to verify the ccrea performance. if the qc result was out of range, new crea a and crea b membranes had to be installed.Radiometer has now resolved the problem (starting with lot 557 and onwards) with narrowed specifications for the membrane material to reduce the diffusion rate through the membrane and has verified that the change brings the performance (drift and lifetime) back to the previous level.
  • 조치
    For membrane units below Lot 557: Continue to perform one quality control measurement every 24 hours using S7835 AutoCheck 6+, level 1 to verify the cCrea performance. If the QC result is out of range, new Crea A and Crea B membranes must be installed. Alternatively, if you wish to return any remaining unused units of cCrea membrane from Lots below 557 please contact your local Radiometer Account Manager on 1800 247 254. For membrane units of Lot 557 onwards: You may return to your standard practice for running QC for the cCrea membranes. This action has been closed-out on 17/08/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    cCrea A+B Membranes used with ABL800 series blood gas analyser rangeItem # 942-073, 08088 Crea A + 08089 Crea B membranLot Numbers: 557 onwardsARTG Number: 229300