Ceiling mounted rail systems for Liko Overhead Lifts 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Hill-Rom Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-01330-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2014-01-16
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The issue relates to fasteners securing overhead lift rail systems to ceilings in chlorinated indoor pool applications. two complaints with no confirmed injuries have been reported to hill-rom that corrosion of the fasteners could lead to the rails becoming unattached and falling. hill-rom’s analysis has concluded that the root cause is stress corrosion cracking (scc) on the stainless steel hardware may occur in the presence of chlorinated environments such as swimming pool areas. there is a risk that continued use of the stationary lift in a chlorinated pool area could result in injuries to patients, caregivers or bystanders.
  • 조치
    Hill-Rom is advising users that an inspection will be carried out on all potentially affected systems located in a chlorinated environment and if systems are found to be of stainless steel hardware, it will be replaced with unaffected hardware. Hill-Rom is reminding users that periodic inspections are required for the rail systems in accordance with the Instructions for Use (IFU). For more details, please see http://www.tga.gov.au/safety/alerts-device-liko-overhead-lifts-140121.htm . This action has been closed-out on 11/02/2016.

Device